Spanish Pressure Monitoring Course

  • Deltex Medical
  • purpose statement
  • Why ODM+ Monitors Arterial Pressure

  • Using the ODM+ with an Arterial Line

    Welcome to Deltex Medical's guide to arterial pressure monitoring with the ODM+.

    The course objectives will give you an understanding of:

    • How to set up arterial pressure monitoring on the ODM+
    • Calibration for Cardiac Output (CO)
    • When to use Doppler flow mode and when to use pressure mode

    This course is for information only and is in no way intended to be a replacement for the Instructions for Use and the Operating Handbook, which should be referred to for full instructions.

    Note: Local laws apply in all cases. The products shown in this course may not be available in all markets.

  • Flow and Pressure on the ODM+

    The ODM+ brings together the minimally invasive oesophageal Doppler monitoring of blood flow with a pulse pressure wave analysis (PPWA) system.

    Oesophageal Doppler is the gold standard for measuring change in cardiac flow and making clinical decisions.

    It is not necessary to have an arterial line to monitor cardiac function using Doppler, however, easy calibration of the ODM+ arterial signal provides continuous monitoring of Cardiac Output (CO). This may be useful if the Doppler signal is unavailable during electrocautery.

    ODM+ also offers the combination parameters of Cardiac Power Output (CPO) and Cardiac Power Index (CPI)

    pressure–flow
  • Blood Flow with Arterial Pressure

    blue–man
  • Using the ODM+ with an Arterial Line

    Arteral line socket

    Connect your existing patient monitor to the ODM+ via the ADC jack plug on the back.

    This feeds your existing arterial signal into the ODM+ to provide pressure information.

    Once you have an optimized Doppler signal, arterial pressure can be calibrated at the touch of a button.

    See the transfer for ODM+

  • The Arterial Waveform

    Pressure measurements taken further away from the aorta are characterised by:

    • A taller systolic peak
    • A deeper dichrotic notch

    On the ODM+ monitor the arterial wave is scaled to an appropriate size for comparison with the Doppler waveform.

    flow-press
    blue-man
  • Using Flow or Pressure Monitoring Mode

    Click to enlarge

    Use Flow Monitoring Mode for direct measurement of flow and to guide interventions.

    Use Pressure Monitoring Mode:

    • if the Doppler signal is unavailable e.g., during electrocautery
    • for continuous monitoring
    • for post-operative monitoring of an awake patient

    If pressure monitoring continues after the Doppler probe has been removed, you will be prompted to recalibrate if the signal deviates by 20% or more. It is then a clinical decision whether to use an ‘awake probe’ (the softer more flexible I2C) for calibration or to sedate the patient and use a standard probe.

    recalibrate
  • Pressure Monitoring Mode

    Calibration – at the touch of a button, needs to be carried out before pressure parameters can be used and should be repeated:

    • When continuous monitoring is likely to be required
    • After a known or suspected change in arterial compliance e.g.:
      • Anaesthesia
      • Other vasoactive drugs
      • Physiological responses to stimuli
    • When there is uncertainty in the arterial pressure-based CO displayed on the ODM+
    • For a calibration alert triggered by 20% variation in pressure-derived SV
    • When it is 12 hours since last calibration and arterial signal will otherwise be terminated
  • How Deltex Medical Derives CO from Arterial Pressure

    Research and development for the ODM+ evaluated 9 available algorithms that convert pressure to flow parameters.

    The Liljestrand and Zander method was chosen because it proved superior to the other algorithms, especially during changes in compliance.1-3

    1 Caillard et al. BJA 2015;114:893-900.
    2 Monge Garcia et al. Critical Care 2013;17:R113.
    3 Sun et al. Crit Care Med 2009;37:72-80.

  • When do I Calibrate

    Any known or suspected change in arterial compliance is an indicator for recalibration.

    compliance

    Arterial compliance is an index of elasticity:
    – calculated from change in blood volume for a given change in blood pressure (BP).

    • Changes in arterial compliance can affect the amplitude and shape of the arterial pressure waveform.
    • Vessels become less compliant (elastic) with increasing age
    • Acute changes in arterial compliance commonly occur during surgery due to interventions such as:
      • Anaesthesia
      • Other vasoactive drugs
      • Physiological responses to stimuli
  • Pressure Monitoring

    art_line

    When any arterial line is used to measure pressure there are three factors that influence the accuracy of the signal:

    • Levelling
    • Zeroing
    • Damp testing

    These should be done before using the arterial signal on the ODM+ and should be checked whenever there is doubt about the arterial signal.

  • Levelling

    Influences accuracy of all arterial lines

    The arterial cannula forms a closed fluid system with the body where we are measuring the force (or height) of a column of water (the tubing) above the transducer. This measurement changes with positioning.

    The right atrium is a suitable reference for levelling the transducer. This is located approximately at the intersection of where a line from the fourth intercostal space at the right side of the sternum (A) meets the midaxillary line (B) – the phlebostatic axis.

    • Invalid if the patient has a barrel shaped chest or severe chest deformity.
    phlebostatic axis

    A consistent angle is best for measuring pressure trends.

    This measurement can be taken with the patient supine or raised to a 60˚ angle.

    Note: changes in the angle of the bed after levelling may give a false result.

    transducer height
    Repeat levelling:
    • At set up of arterial line and prior to using ODM+
    • At handover
    • When any doubt exists about the arterial readings
    • Whenever patient position is altered, the transducer height should be altered.
  • Zeroing

    Influences accuracy of all arterial lines

    transducer

    Atmospheric pressure is a constant in the room. Just as weighing scales can be set to zero, atmospheric pressure can be eliminated from the readings by exposing the column of fluid to the pressure in the room. This requires the 3-way tap to be closed to the patient. A value of ‘zero’ can then be set on the monitor.

    This is usually performed:

    • At set up of arterial line and prior to using ODM+
    • Once per shift, or at handover
    • After being disconnected from patient’s main monitor
    • When any doubt exists about the readings
  • Damp Testing (Square Test)

    Influences accuracy of all arterial lines

    This test is designed to evaluate vibration detected by the transducer and the degree to which it is restrained or damped.

    When the flush valve is opened and fluid is pushed quickly through the giving set, a square wave is created in response on the arterial trace on the patient’s main monitor. The degree of oscillation detected by the transducer after the flush reflects the degree of damping in the system.

    Square test
  • Damping – Too Much or Too Little?

    Too much variation is observed when more than two oscillations occur before returning to baseline.

    The system is underdamped which may be caused by:

    • Catheter whip or artefact
    • Stiff non-compliant tubing
    • Hypothermia
    • Tachycardia or dysrhythmia

    underdamped wave

    Importantly, systole would appear falsely high and diastole falsely low.

    If there are no oscillations after the square wave, or the response is sluggish, the system is overdamped. Causes can be:

    • Loose connections
    • Air bubbles
    • Kinks
    • Blood clots
    • Arterial spasm
    • Narrow tubing

    overdamped wave

    Importantly, systolic values can be underestimated and diastolic overestimated.

    Damp testing should take place at:
    • At set up of arterial line and prior to using ODM+
    • At handover
    • Whenever you are unsure of the arterial readings
    • If the arterial line is disconnected from the patient monitor.
  • Parameters

    Click parameter to see definition

    Click on abbreviation for full term.

    The data boxes can be customized when switching the monitor on by pressing ‘Monitor settings’ and then ‘User settings’. Use the large ‘Control knob’ to identify and select the parameter for each box.

  • Thank You for Joining E-learning with Deltex Medical

    Further modules are to be released soon.

    • Practical application of the Oesophageal Doppler Simulator
    • Accuracy and precision
    • Clinical waveforms

Posted in English